6 Mounting
FLUXUS H721 6.3 Temperature probe
65
UMFLUXUS_H721V1-5EN, 2022-05-15
6.3.2.3 Installation with quick release clasp
• Cut the tension strap to length (pipe circumference + at least 120 mm).
• Position the temperature probe on the pipe (see Fig. 6.78).
• Place the tension strap around the temperature probe and the pipe.
• Insert the tension strap into the clasp.
• Tighten the tension strap.
• Tighten the screw of the clasp.
The edge of the tension strap is very sharp.
Risk of injury!
→ Debur sharp edges.
→ Wear the required personal protective equipment.
→ Observe the applicable rules.
Fig. 6.80: Quick release clasp