13. Specifications
13.1 General specifications
Display counts: 0~6000
Measuring rate: 3 times per second
Over-range indication: OL or –OL
Auto power off: Programmable: OFF, 1, 2, 5, or 10 (default) minutes
Low battery indicators: and low battery informaton screens appear
Power requirement: 3 × 1.5 V ‘AA’ alkaline/lithium batteries or TA04 rechargeable battery
Approximate battery life for thermal imager:
● 2.5 hours: Alkaline ‘AA’ Battery x 3
● 12 hours: Energizer L91 Lithium (Li/FeS
2
) ‘AA’ Battery x 3
● 12 hours: Optional Rechargeable Battery: Li-Polymer; FLIR PN: TA04-KIT
Calibration: 1-year calibration cycle
Operating temperature:
32 to 86°F (0 to 30°C) (≤ 80% RH)
86 to 104°F (30 to 40°C) (≤ 75% RH)
104 to 122°F (40 to 50°C) (≤ 45%RH)
Storage temperature:
4 to 140°F (-20 to 60°C)
Relative Humidity: 0~80% RH (batteries not installed)
Temperature coefficient: 0.2 × (specified accuracy)/°C, <64.4°F (18°C), >82.4°F (28°C)
Operating altitude: 6562ʹ (2000m)
Jaw opening: 1.38in (35mm)
Pollution degree: 2
Dimensions: (D × W × L): 1.91 × 3.82 × 10.04 in (48.5 × 97 × 255 mm)
Weight: 16.2 oz (460g) without batteries
Agency Approvals: UL, CE, RCM
Over-voltage category: EN 61010-1 CAT IV-600 V, CAT III-1000 V, EN 61010-2-032
Installation sources, utility transformers, all outside conductors, counters, protective
devices on primary sides and electricity meters