Fluke 5080A - AC Current (Sine Wave)

Fluke 5080A
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5080A
Service Manual
10
AC Current (Sine Wave)
Range Frequency
Specification, tcal ±5 °C
±(% of output + μA)
Compliance
Adder
[2]
(μA/V)
Max. Distortion &
Noise 10 Hz to 10 kHz
Bandwidth ±(% of
output + floor)
Max.
Inductive
Load (μH)
90 days 1 year
LCOMP OFF
29.0 to
329.9 μA
45 to 65 Hz 0.24 % + 0.75 0.25 % + 0.75 0.05
0.2 % + 3 μA
200
65 Hz to 1 kHz 0.25 % + 0.75 0.26 % + 0.75 0.15
0.33 to
3.2999 mA
45 to 65 Hz 0.21 % + 0.9 0.22 % + 0.9 0.05
0.2 % + 5 μA
200
65 Hz to 1 kHz 0.22 % + 0.9 0.23 % + 0.9 0.15
3.3 to
32.999 mA
45 to 65 Hz 0.09 % + 12 0.10 % + 12 0.05
0.2 % + 15 μA
50
65 Hz to 1 kHz 0.18 % + 12 0.19 % + 12 0.15
33 to
329.99 mA
45 to 65 Hz 0.09 % + 120 0.10 % +120 0.1
0.2 % + 150 μA
50
65 Hz to 1 kHz 0.18 % + 120 0.19 % +120 0.2
0.33 to
1.0999 A
45 to 65 Hz 0.09 % + 1200 0.10 % + 1200 10
0.35 % + 1.5 mA 2.5
65 Hz to 1 kHz 0.22 % + 1200 0.24 % + 1200 125
1.1 to
2.9999 A
45 to 65 Hz 0.09 % + 1500 0.10 % + 1500 10
0.35 % + 1.5 mA 2.5
65 Hz to 1 kHz 0.26 % + 1500 0.28 % + 1500 125
3.0 to
10.999 A
45 to 65 Hz 0.24 % + 6000 0.25 % + 6000 10
0.6 % + 15 mA 1
65 Hz to 1 kHz 0.38 % + 6000 0.40 % + 6000 125
11 to
20.500 A
[1]
45 to 65 Hz 0.48 % + 15000 0.50 % + 15000 10
0.6 % + 15 mA 1
65 Hz to 1 kHz 0.50 % + 15000 0.52 % + 15000 125
LCOMP ON
29.0 to
329.9 μA
45 to 65 Hz
0.24 % + 0.75 0.25 % + 0.75 0.05
0.3 % + 3 μA
2.5 H
[3]
0.33 to
3.2999 mA
0.21 % + 0.9 0.22 % + 0.9 0.05
0.5 % + 5 μA
3.3 to
32.999 mA
0.19 % + 9 0.20 % + 9 0.05
0.5 % + 15 μA
33 to
329.99 mA
0.19 % + 90 0.20 % + 90 0.1
0.5 % + 150 μA
0.33 to
1.0999 A
0.20 % + 900 0.21 % + 900 10 0.6 % + 1.5 mA
1.1 to
2.9999 A
0.22 % + 900 0.23 % + 900 10 0.6 % + 1.5 mA
3.0 to
10.999 A
0.24 % + 6000 0.25 % + 6000 10 0.6 % + 1.5 mA
11 to
20.500 A
[1]
0.48 % + 15000 0.50 % + 15000 10 0.6 % + 1.5 mA
[1] Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, the current may be provided 60 T-I minutes in any
60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in amps. For
example, 17 A at 23 °C could be provided for 60-17-23 = 20 minutes each hour. When the 5080A is outputting currents between 5
and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those conditions, the allowable "on" time
indicated by the formula is achieved only after the 5080A is outputting currents <5 A for the "off" period first.
[2] To be applied for compliance voltages >1 V rms.
[3] Subject to compliance voltage limits.
Range
Resolution (μA)
Max. Compliance Voltage,
LCOMP Off, V rms
Max. Compliance Voltage,
LCOMP On, V rms
29.0 to 329.9 μA
0.1
3.3
[1]
3.3
[1]
0.33 to 3.2999 mA 0.1 6.5 6.5
3.3 to 32.999 mA 1 6.5 44
33 to 329.99 mA 10 6 25
0.33 to 2.9999 A 100 4 4
3 to 20.500 A 1000 3 3
[1] Load impedance <10 kΩ.