Calibrate Through the Remote Interface
To access the standard calibration steps through the remote interface, send the
command:
CAL_START MAIN
To jump to a specified calibration step, add a comma followed by a modifier to
specify an entry point. Table 50 shows the applicable entry points and the
modifier to use for each point.
Table 50. Jumping to a Specific Calibration Step in Remote
Entry points for CAL_START MAIN Modifier
AC Volts AV
Thermocouple Measuring TEMPX
DC Current ICAL
AC Current AI
AUX DC Volts V2
AUX AC Volts AVS
Resistance R
Capacitance C
Entry Points for CAL_START FACTORY Modifier
NORMAL Volts and AUX Volts Phase PHASE
Volts and Current Phase IPHASE
For example, to jump directly to ac volts calibration, send the command:
CAL_START MAIN,AV
To go directly to Resistance calibration, send the command:
CAL_START MAIN,R
These calibration commands can be sent through the IEEE-488 or serial
interface. To type in commands through the serial interface:
1. Use a Fluke PM8914 cable to connect the applicable COM port on a PC to
the 5080A Serial 1 connector.
2. In Microsoft Windows, start the Terminal program. Set the communications
parameters to agree with that of the Product.
3. Push . At the prompt, type the calibration command, for example
CAL_START MAIN.