Calibrator
Calibrate Through the Remote Interface
75
Remote Commands
The IEEE-488/RS-232 remote calibration commands for the Product are shown
in this section alphabetically. This list ignores the * character that is used as the
first character in the common commands. The remote commands duplicate
works that can be done from the front panel in local operation.
Note
For remote commands pertaining to normal operation of the Product,
see the Operators Manual.
IEEE-488 (GPIB) and RS-232 Applicability Each command title shown in this
section shares the same remote interface applicability, IEEE-488 (general
purpose interface bus, or GPIB) and RS-232 remote operations, and command
group: Sequential, Overlapped, and Coupled.
Sequential Commands – Commands executed immediately as they are found
in the data stream are sequential commands. All commands that is not
overlapped or coupled are sequential.
Overlapped Commands – Commands that take added time to execute are
overlapped commands. The execution time of these commands can overlap the
subsequent command.
Coupled Commands – Commands that “couple” in a compound command
sequence are coupled commands. Some examples are CUR_POST and OUT.
You must take precautions to be sure one command does not disable a second
command and cause a fault.
CAL_ABORT
Description Instructs the Product to abort a calibration procedure after current
step.
Examples CAL_ABORT
CAL_BACKUP
Description Skip to a previous entry point in the current calibration procedure.
Examples CAL_BACKUP
CAL_DATE?
Description Returns a calibration date related to stored calibration constants.
The date is returned with the same format as the CLOCK command.
Parameters MAIN Returns the cal date of the main unit calibration.
ZERO Returns the cal date of the ZERO calibration.
OHMSZERO Returns the cal date of the Ohms zero calibration.
SCOPE Returns the cal date of the scope option calibration.
Example CAL_DATE? MAIN
Response The date.

Table of Contents

Other manuals for Fluke 5080A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5080A and is the answer not in the manual?

Fluke 5080A Specifications

General IconGeneral
BrandFluke
Model5080A
CategoryTest Equipment
LanguageEnglish

Related product manuals