5080A
Service Manual
76
CAL_DAYS?
Description Gets the number of days and hours since the last calibration
constant. The data output for date is in the same format as the
CLOCK command.
Parameters MAIN Returns the days since the main unit calibration.
ZERO Returns the days since the ZERO calibration.
OHMSZERO Returns the days since the Ohms zero calibration.
SCOPE Returns the days since the scope option calibration.
Example CAL_DAYS? ZERO
Response (Integer) Days.
(Integer) Hours.
CAL_FACT
Description Set the procedure “fault action” flag. Procedures refer to calibration
and diagnostic procedures. This command is more useful for
diagnostics that calibration.
Parameters CONT To continue on faults.
ABORT To abort on faults.
Example CAL_FACT ABORT
CAL_FACT?
Description Get the procedure “fault action” flag.
Example CAL_FACT?
Response ABORT.
CAL_FAULT?
Description Get information about calibration error, if one occurred).
Example CAL_FAULT?
Response 1. Error number (use EXPLAIN? command to interpret).
2. Name of step where error occured.
CAL_INFO?
Description Return message or instructions related to the running step.
Example CAL_INFO?
Response (String) the message string.
CAL_NEXT
Description Continue a calibration procedure if it is waiting for a CAL_NEXT
command.
Parameters (Optional) reference value (used if it’s waiting for a reference). If the
reference value has no unit, the unit is assumed to be that returned
by the CAL_REF? command.
Example CAL_NEXT