5080A
Service Manual
46
Fluke 8508A
Set the Fluke 8508A
to external guard
UUT
gru005.eps
Figure 10. DC Voltage Adjustment Setup
AC Volts Calibration (NORMAL Output)
The equipment shown in Table 32 is necessary for calibration of the ac volts
function. (The equipment is also listed in the consolidated table, Table 29.)
Table 32. Test Equipment Required for AC Volts Calibration
Quantity Manufacturer Model Equipment
1 Fluke PN 900394 Type N to dual banana adapter
1 Fluke 5790A, 5790B AC Measurement Standard
To calibrate the ac voltage function:
1. Measure the Product output through Input 1 of a Fluke 5790A or 5790B AC
Measurement Standard. Use a Type N to dual banana adapter as shown in
Figure 11..
2. Type in the measured values into the Product for each step in Table 33 as
prompted.
AC Voltage adjust completed.