Calibrator
SC200 Calibration
57
Fluke 8508A
UUT
Guard to Normal Lo Terminal
gru009.eps
Figure 16. Connections for Calibrating 2-Wire Resistance
SC200 Calibration
The SC200 is an oscilloscope calibration option for the 5080A. The sections that
follow explain how to calibrate the SC200 Oscilloscope Option.
Note
Although this procedure will calibrate the SC200 option in the field, it
is recommended the Product be sent to Fluke Calibration for
calibration and verification.
Fluke Calibration recommends that you review all of the procedures in this
section to make sure you have the resources to complete the calibration before
you start.
Required Test Equipment
All test equipment used to calibrate the SC200 option must be calibrated and in
their specified environment. To keep traceability, all equipment must be certified
traceable. Refer to the operators manual of each piece of test equipment for
operation instructions.
Table 42 shows the test equipment required to verify and calibrate the SC200
option.
Table 42. Required Test Equipment for SC200 Calibration
Test Equipment Recommended Model Minimum Specifications
Wave Generator, Edge Amplitude Calibration, AC Voltage Verification
Digital Multimeter Hewlett Packard 3458A Voltage 1.8 mv to ±105 V p-p
uncertainty: 0.06 %
Edge 4.5 mV to 2.75 V p-p
uncertainty: 0.06 %
Adapter Pomona #1269 BNC(f) to Double Banana Plug
Termination Feedthrough 50 Ω, +1 % (used with Edge
Amplitude Calibration and AC Voltage
Verification)
BNC Cable (supplied with SC200)