Service Manual
56
Table 41. Calibration Steps for Resistance
Step Output (4-Wire Ohms, NORMAL and AUX)
1 1.0000 Ω
2 1.9000 Ω
3 10.000 Ω
4 19.000 Ω
5 100.000 Ω
6 190.000 Ω
7 1.00000 kΩ
8 1.90000 kΩ
9 10.0000 kΩ
10 19.0000 kΩ
11 100.000 kΩ
12 190.000 kΩ
2-Wire Ohms, NORMAL
13 1.00000 MΩ
14 1.90000 MΩ
15 10.0000 MΩ
16 19.000 MΩ
17 100.00 MΩ
18 190.00 MΩ
Fluke 8508A
Sense Input
Guard to Normal Lo Terminal
UUT
gru008.epx
Figure 15. Connections for Calibrating 4-Wire Resistance