The Performance Tests section contains the procedures used to make sure the
Product performs to specifications.
Required Test Equipment
Table 3 is a list of the test equipment necessary to perform Verification and
Calibration of the main Product functions.
Table 3. Test Equipment for Verification and Calibration
Test Equipment Recommended Model
Reference Multimeter Fluke 8508A
AC Measurement Standard
Fluke 5790A, Fluke 5790B, or equivalent.
Note
The 5790A and 5790B have
identical connections. This manual
shows the 5790A in all of the
illustrations but also apply to the
5790B.
Current shunts (10 mA, 50 mA, 20 A, 1 A, 2 A,
and 200 mA)
Fluke A40B
Shunt, 0.01 Ω Guildline 9230
Resistance Standard, 1 kΩ Fluke 742A-1k
Resistance Standard, 100 Ω Fluke 742A-100
Resistance Standard, 10 Ω Fluke 742A-10
Resistance Standard, 1 Ω Fluke 742A-1
Type-N Dual Banana Adapter Fluke Part Number 900394
A40 Current Shunt Fluke 792-7004
Distortion Analyzer Krohn-Hite 6900B
Phasemeter Clarke-Hess 6000
Timer/Counter Fluke PM6680B
Verify the Main Calibrator
Use Tables 4 through 14 to make sure the Product operates to its specifications.
The tables are for approved metrology personnel who have access to a
standards laboratory that has equipment that can test calibration equipment of
this level of accuracy. The tables show the recommended test points and the
upper and lower limits for each point. The limits calculation is the 90-day
specification added to or subtract from the output value. There is no built-in factor
for measurement uncertainty.