Calibrator
Mainframe Calibration
45
DC Volts Calibration (NORMAL Output)
The equipment shown in Table 30 is necessary for calibration of the dc volts
function. (The equipment is also listed in the consolidated table, Table 29.)
Table 30. Test Equipment Required for DC Volts Calibration
Quantity Manufacturer Model Equipment
1 Fluke 8508A Reference Multimeter
To calibrate the dc voltage function:
1. Make sure that the UUT is in Standby.
2. Start calibration as instructed in the Start Calibration section.
3. Perform an internal DC Zeros Calibration as prompted.
4. Connect the test equipment as shown in Figure 10.
5. Measure and enter the values from the UUT in Table 31 as prompted by the
Product. Disconnect and reconnect the DMM as prompted during these
steps.
DC Voltage adjust completed.
Table 31. Calibration Steps for DC Volts
Step 5080A Output (NORMAL)
1 1.000000 V
2 3.000000 V
3 -1.000000 V
4 -3.000000 V
5 0.0000 mV
6 300.0000 mV
7 30.00000 V
8 300.0000 V
9 1000.000 V