Service Manual
24
Table 4. Verification Tests for DC Voltage (Normal)
Range Output Lower Limit Upper Limit
329.999 mV 0.000 mV -0.0100 mV 0.0100 mV
329.999 mV 329.000 mV 328.9538 mV 329.0462 mV
329.999 mV -329.000 mV -329.0462 mV -328.9538 mV
3.29999 V 0.000000 V -0.000015 V 0.000015 V
3.29999 V 1.00000 V 0. 999905 V 1.000095 V
3.29999 V -1.00000 V -1.000095 V --0.999905 V
3.29999 V 3.29000 V 3.289722 V 3.290278 V
3.29999 V -3.29000 V -3.290278 V -3.289722 V
32.9999 V 0.00000 V -0.00015 V 0.00015 V
32.9999 V 10.0000 V 9.99905 V 10.00095 V
32.9999 V -10.0000 V -10.00095 V -9.99905 V
32.9999 V 32.9000 V 32.89722 V 32.90278 V
32.9999 V -32.9000 V -32.90278 V -32.89722 V
100.000 V 100.000 V 99.9885 V 100.0115 V
100.000 V -100.000 V -100.0115 V -99.9885 V
329.999 V 329.000 V 328.9656 V 329.0344 V
329.999 V -329.000 V -329.0344 V -328.9656 V
1000.00 V 330.000 V 329.962 V 328.9656 V
1000.00 V -330.000 V -330.039 V -329.962 V
1000.00 V 1020.00 V 1019.893 V 1020.108 V
1000.00 V -1020.00 V -1020.108 V -1019.893 V
Table 5. Verification Tests for DC Voltage (AUX)
Range Output Lower Limit Upper Limit
329.999 mV 0.00 mV -1.000 mV 1.000 mV
329.999 mV 329.00 mV 327.671 mV 330.329 mV
329.999 mV -329.00 mV -330.329 mV -327.671 mV
3.29999 V 3.2900 V 3.28571 V 3.29429 V
3.29999 V -3.2900 V -3.29429 V -3.28571 V
7.0000 V 7.000 V 6.9920 V 7.0080 V
7.0000 V -7.000 V -7.0080 V -6.9920 V