EasyManua.ls Logo

Fluke 5080A - Page 35

Fluke 5080A
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrator
Performance Tests
25
Table 6. Verification Tests for DC Current
Range Output Lower Limit Upper Limit
330 μA 0.00 μA -0.100 μA 0.100 μA
330 μA 329.00 μA 328.670 μA 329.330 μA
330 μA -329.00 μA -329.330 μA -328.670 μA
3.3 mA 0.00000 mA -0.000250 mA 0.000250 mA
3.3 mA 3.2900 mA 3.28778 mA 3.29222 mA
3.3 mA -3.00000 mA -3.29222 mA -3.28778 mA
33 mA 0.00000 mA -0.001250 mA 0.001250 mA
33 mA 32.900 mA 32.8830 mA 32.9170 mA
33 mA -32.900 mA -32.9170 mA -32.8830 mA
330 mA 0.00000 mA -0.016500 mA 0.016500 mA
330 mA 329.00 mA 328.826 mA 329.174 mA
330 mA -329.00 mA -329.174 mA -328.826 mA
1.1 A 0.00000 A -0.000220 A 0.000220 A
1.1 A 1.0000 A 0.99838 A 1.00162 A
1.1 A -1.0000 A -1.00162 A -0.99838 A
3 A 0.00000 A -0.000220 A 0.000220 A
3 A 2.9900 A 2.98440 A 2.99560 A
3 A -2.9900 A -2.99560 A -2.98440 A
11 A 0.00000 A -0.002500 A 0.002500 A
11 A 10.900 A 10.8724 A 10.9276 A
11 A -10.900 A -10.9276 A -10.8724 A
20.5 A 0.00000 A -0.003750 0.003750 A
20.5 A 19.9900 A 19.8007 A 19.9993 A
20.5 A -19.9900 A -19.9993 A -19.8007 A