Table 7. Verification Tests for 4-Wire Resistance
Range Output Tolerance
0.0000 Ω 0.0000 Ω ±0.01000 Ω
1.0000 Ω 1.0000 Ω ±0.00989 Ω
1.9000 Ω 1.9000 Ω ±0.00931 Ω
10.000 Ω 10.000 Ω ±0.014 Ω
19.000 Ω 19.000 Ω ±0.0171 Ω
100.000 Ω 100.000 Ω ±0.035 Ω
190.000 Ω 190.000 Ω ±0.0665 Ω
1.00000 kΩ 1.00000 kΩ ±0.00022 kΩ
1.90000 kΩ 1.90000 kΩ ±0.000418 kΩ
10.0000 kΩ 10.0000 kΩ ±0.0022 kΩ
19.0000 kΩ 19.0000 kΩ ±0.00494 kΩ
100.000 kΩ 100.000 kΩ ±0.035 kΩ
190.000 kΩ 190.000 kΩ ±0.0741 kΩ
1.00000 MΩ 1.00000 MΩ ±0.00035 MΩ
1.90000 MΩ 1.90000 MΩ ±0.000665 MΩ
10.000 MΩ 10.000 MΩ ±0.0092 MΩ
19.000 MΩ 19.000 MΩ ±0.0266 MΩ
100.00 MΩ 100.00 MΩ ±0.049 MΩ
190.00 MΩ 190.00 MΩ ±1.876 MΩ
Table 8. Verification Tests for AC Voltage (Normal)
Range Output Frequency Lower Limit Upper Limit
33 mV 10.00 mV 45 Hz 9.909 mV 10.091 mV
33 mV 10.00 mV 100 Hz 9.908 mV 10.092 mV
33 mV 10.00 mV 1 kHz 9.908 mV 10.092 mV
33 mV 32.90 mV 45 Hz 32.738 V 30.062 V
33 mV 32.90 mV 100 Hz 32.735 V 33.065 V
33 mV 32.90 mV 1 kHz 32.735 V 33.065 V
330 mV 329.00 mV 45 Hz 328.512 V 329.488 V
330 mV 329.00 mV 100 Hz 328.479 V 329.521 V
330 mV 329.00 mV 1 kHz 328.479 V 329.521 V
3.3 V 3.2900 V 45 Hz 3.28686 V 3.29314V
3.3 V 3.2900 V 100 Hz 3.28653 V 3.29347 V