5080A
Service Manual
58
Table 42. Required Test Equipment for SC200 Calibration (cont.)
Test Equipment Recommended Model Minimum Specifications
Edge Rise Time and Aberrations Verification
High Frequency
Digital Storage
Oscilloscope
Tektronix 11801 with SC-
22/26 sampling head, or
Tektronix TDX 820 with
8 GHz bandwidth
Frequency 2 GHz
Resolution 4.5 mV to 2.75 V
Attenuator Weinschel 9-10 (SMA) or
Weinschel 18W-10 or
equivalent
10 dB, 3.5 mm(m/f)
Adapter BNC(f) to 3.5 mm(m)
BNC Cable (supplied with SC200)
Leveled Sine Wave Amplitude Calibration and Verification
AC Measurement
Standard
Fluke 5790A
Range 5 mV p-p to 5.5 V p-p
Frequency 50 kHz
Adapter Pomona #1269 BNC(f) to Double Banana Plug
Termination Feedthrough 50 Ω, +1 %
BNC Cable (supplied with SC200)
DC and AC Voltage Calibration and Verification, DC Voltage Verification
Digital Multimeter Hewlett Packard 3458A
Adapter Pomona #1269 BNC(f) to Double Banana Plug
Termination Feedthrough 50 Ω, +1 %
BNC Cable (supplied with SC200)
Leveled Sine Wave Frequency Verification
Frequency Counter
PM 6680 with Option (PM
9621, PM 9624, or
PM 9625) and (PM 9678)
50 kHz to 350 MHz, <1.6 ppm uncertainty
Adapter Pomona #3288 BNC(f) to Type N(m)
BNC Cable (supplied with SC200)
Leveled Sine Wave Flatness (Low Frequency) Calibration and Verification
AC Measurement
Standard
Fluke 5790A with 03
option
Range 5 mV p-p to 5.5 V p-p
Frequency 50 kHz to 10 MHz
Adapter Pomona #3288 BNC(f) to Type N(m)
BNC Cable (supplied with SC200)
Leveled Sine Wave Harmonics Verification
Spectrum Analyzer Hewlett Packard 8590A
Adapter Pomona #3288 BNC(f) to Type N(m)
BNC Cable (supplied with SC200)