Calibrator
Mainframe Calibration
47
UUT5790A or 5790B
5790A
AC MEASUREMENT
STANDARD
1000V RMS MAX 1000V RMS MAX
3V RMS MAX
INPUT 1 INPUT 2
SHUNT
HI
LO
GUARDGROUND
SHELL FLOATING
10V PEAK
MAX
WIDEBAND
7V RMS MAX
MAX
10V PK
SHELL FLOATING
Set the 5790A or 5790B
to external guard
gru104.eps
Figure 11. AC Voltage Adjustment Setup
Table 33. Calibration Steps for AC Volts
Steps
Output (NORMAL)
Amplitude Frequency
1 3.2999 V 100.00 Hz
2 0.3300 V 100.00 Hz
3 30.000 mV 100.00 Hz
4 300.000 mV 100.00 Hz
5 30.0000 V 100.00 Hz
6 300.000 V 100.00 Hz
7 1000.00 V 100.00 Hz