Calibrator
Calibrate Through the Remote Interface
77
CAL_REF?
Description Return normal value expected for reference entry.
Response 1. The nominal value.
2. The accepted or implied unit.
Example 3.000000e+00,V
CAL_SKIP
Description Skip to subsequent entry point in calibration procedure.
Examples CAL_SKIP
CAL_SECT
Description Skip to next section of the calibration procedure.
Example CAL_SECT
CAL_START
Description Start a calibration procedure.
Parameters MAIN Procedure for the 5080A minus the scope cal option.
ZERO Internal procedure to touch up zero offsets.
OHMSZERO Internal procedure to touch up resistance offsets.
SCOPE Procedure for the 5080A-SC200 scope cal option.
DIAG Diagnostic pseudo-cal procedure.
NOT Aborts a procedure after the step underway.
A second optional parameter that shows the name of the step at
which to start can be used. If this parameter is not supplied,
calibration starts at the first calibration step.
Examples CAL_START MAIN Start the MAIN calibration
procedure.
CAL_START MAIN,DVG3_3 Start the MAIN calibration procedure
at DVG3_3 point.
CAL_STATE?
Description Gets the state of calibration.
Response RUN A calibration step is in progress.
REF Waiting for a CAL_NEXT with reference (measurement)
value.
INS Instruction available, waiting for a CAL_NEXT.
NOT Not in a calibration procedure (or at end of one)
CAL_STEP?
Description Gets the name of the calibration step in progress.
Response (Char) the step name.
Examples IDAC_RATIO (running IDAC ratio calibration)
NOT (not running a calibration procedure now)
CAL_STORE
Description Store new calibration constants (CAL switch must be ENABLED).
Example CAL_STORE