5080A
Service Manual
78
CAL_STORE?
Description Gets if a CAL_STORE is necessary.
Response 1 Yes.
0 No
Example CAL_STORE?
CAL_SW?
Description Gets the position of the calibration switch.
Response 1 Enabled.
0 Normal
Example CAL_SW?
EOFSTR
Description Sets the End-Of-File character string used for calibration reports.
The maximum length is two characters. The EOF parameter is kept
in nonvolatile memory.
Parameters The EOF string (two characters maximum).
EOFSTR?
Description Gets the End-Of-File character string used for calibration reports.
Parameters None
Response (String) The End-Of-File character string.
PR_RPT
Description Prints a self-calibration report out the selected serial port.
Parameters 1. Type of report to print: STORED, ACTIVE, or CONSTS
2. Format of report: PRINT (designed to be read) SPREAD
(designed to be loaded into a spreadsheet)
3. Calibration interval to be used for instrument specifications in the
report: I90D (90 day specifications) or I1Y (1 year specifications)
4. Serial port out which to print report: HOST or UUT
Example PR_RPT STORED,PRINT,I90D,HOST
RPT?
Description Gets a self-calibration report.
Parameters 1. Type of report to print: STORED, ACTIVE, or CONSTS
2. Format of report: PRINT (designed to be read) SPREAD
(designed to be loaded into a spreadsheet)
3. Calibration interval to be used for instrument specifications in the
report: I90D (90 day specifications) or I1Y (1 year specifications)
Example PR_RPT? STORED,PRINT,I90D
RPT_PLEN
Description Sets the page length used for calibration reports. This parameter is
kept in nonvolatile memory.
Parameters Page length.