28
Table 10. Verification Tests for AC Current (cont.)
Range Output Frequency Lower Limit Upper Limit
3.3 mA 3.2900 mA 45 Hz 3.28219 mA 3.29781 mA
3.3 mA 3.2900 mA 100 Hz 3.28186 mA 3.29814 mA
3.3 mA 3.2900 mA 1 kHz 3.28186 mA 3.29814 mA
33 mA 19.900 mA 45 Hz 19.8701 mA 19.9299 mA
33 mA 19.900 mA 100 Hz 19.8701 mA 19.9299 mA
33 mA 32.900 mA 1 kHz 32.8288 mA 32.9712 mA
330 mA 199.00 mA 45 Hz 198.701 mA 199.299 mA
330 mA 199.00 mA 100 Hz 198.701 mA 199.299 mA
330 mA 329.00 mA 1 kHz 328.288 mA 329.712 mA
1.1 A 1.0000 A 45 Hz 0.99790 A 1.00210 A
1.1 A 1.0000 A 100 Hz 0.99790 A 1.00210 A
1.1 A 1.0000 A 1 kHz 0.99660 A 1.00340 A
3 A 1.9900 A 45 Hz 1.98671 A 1.99329 A
3 A 1.9900 A 100 Hz 1.98671 A 1.99329 A
3 A 2.9900 A 1 kHz 2.98073 A 2.99927 A
11 A 10.000 A 45 Hz 9.9700 A 10.0300 A
11 A 10.000 A 100 Hz 9.9700 A 10.0300 A
11 A 10.000 A 1 kHz 9.9560 A 10.0440 A
20.5 A 19.900 A 45 Hz 19.7895 A 20.0105 A
20.5 A 19.900 A 100 Hz 19.7855 A 20.0145 A
20.5 A 19.900 A 1 kHz 19.7855 A 20.0145 A
Table 11. Verification Tests for Phase (Normal Voltage vs. AUX Current)
Range,
Normal
Output
Output,
Normal V
Frequency
Range, AUX
Output
Current
Output
AUX
Phase
Lower
Limit
Upper
Limit
3.3 V 3.00000 V 45 Hz 3.3 mA 300.00 mA 0 ° -0.25 ° 0.25 °
3.3 V 3.00000 V 500 Hz 3.3 mA 300.00 mA 0 ° -1.5 ° 1.5 °
3.3 V 3.00000 V 1 kHz 3.3 mA 300.00 mA 0 ° -5 ° 5 °
3.3 V 3.00000 V 65 Hz 2.5 A 2.00000 A 0 ° -0.25 ° 0.25 °
3.3 V 3.00000 V 65 Hz 2.5 A 5.0000 A 0 ° -0.25 ° 0.25 °