5080A
Service Manual
32
Table 16. Megohm Option HVR Verification Point (cont.)
Range Output Tolerance
6.9800 MΩ 6.9800 MΩ ±0.02094 MΩ
10.000 MΩ 10.000 MΩ ±0.0500 MΩ
10.240 MΩ 10.240 MΩ ±0.0512 MΩ
20.980 MΩ 20.980 MΩ ±0.1049 MΩ
39.190 MΩ 39.190 MΩ ±0.196 MΩ
76.550 MΩ 76.550 MΩ ±0.3827 MΩ
100.00 MΩ 100.00 MΩ ±0.500 MΩ
138.60 MΩ 138.60 MΩ ±0.693 MΩ
148.90 MΩ 148.90 MΩ ±0.744 MΩ
289.60 MΩ 289.60 MΩ ±1.448 MΩ
559.60 MΩ 559.60 MΩ ±2.798 MΩ
6. Connect the Product directly to the megohmmeter as shown in Figure 9.
Make sure you reverse the polarity.
Note
When you use the 100 G
Ω
value on the Product for some
megohmmeters, the leads must be swapped between the HI and LO
ohms resistance output on the Product. The ground must be turned
on (Gnd On) when you swap HI and LO leads positions in the high
ohms resistance function. For example, to make a correct
measurement with the Quadtech 1865 Megohmmeter, connect the
HI terminal on the megohmmeter to the LO terminal on the Product
and connect the LO terminal on the megohmmeter to the HI terminal
on the Product. Turn the ground on and make the measurement.
Qu
adtech
1
8
65
+
-
Fluke 5080A
Megohmmeter
gru003.eps
Figure 9. High Resistance Verification for 1 G Ω and higher