5080A
Service Manual
52
Table 37. Calibration Steps for AC Current
Steps
Output (AUX HI, LO)
Amplitude Frequency Shunt to Use
1 3.2999 mA 55.00 Hz Fluke A40 10 mA
2 0.3300 mA 55.00 Hz Fluke A40 10 mA
3 3.0000 mA 55.00Hz Fluke A40 10 mA
4 3.00000 mA 1.000 kHz Fluke A40 10 mA
5 300.0uA 55.00 Hz Fluke A40 10 mA
6 300.0uA 1.000 kHz Fluke A40 10 mA
7 30.000 mA 55.00 Hz Fluke A40 200 mA
8 30.000 mA 1.000 kHz Fluke A40 200 mA
9 300.000 mA 55.00 Hz Fluke A40 2 A
12 300.000 mA 1.000 kHz Fluke A40 2 A
13 1.9000 A 55.00 Hz Fluke A40 2 A
15 1.9000 A 1.000 kHz Fluke A40 2 A
AUX 20A, LO
16 10.0000 A 55.00 Hz Fluke A40A 20 A
17 10.0000 A 1.000 kHz Fluke A40A 20 A