10. If you are using an absolute-pressure type
module (Model Number starts with “700PA”)
zero the pressure module as follows:
a. Vent the module to atmosphere.
b. Press the SET OFFSET softkey.
c. Enter the ambient atmospheric pressure in
the units currently displayed.
Note
Do not rely on airport pressure reports.
Use a barometric pressure standard in the
same area as the Calibrator.