EasyManuals Logo

Fluke 726 User Manual

Fluke 726
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
Multifunction Process Calibrator
Safety Information
5
When using a pressure module, make sure the process pressure line is shut off and depressurized
before connecting it or disconnecting it from the pressure module.
Use only 4 AA batteries, properly installed in the Calibrator case, to power the Calibrator.
Disconnect test leads before changing to another measure or source function.
When servicing the Calibrator, use only specified replacement parts.
To avoid false readings, which could lead to possible electric shock or personal injury, replace the
battery as soon as the battery indicator (M) appears.
Turn off circuit power before connecting the Calibrator mA and COM terminals in the circuit. Place
Calibrator in series with the circuit.
Do not allow water into the case.
WCaution
To avoid possible damage to the Calibrator or to equipment under test:
Disconnect the power and discharge all high-voltage capacitors before testing resistance or continuity.
Use the proper input jacks, function, and range for the measurement or sourcing application.

Table of Contents

Other manuals for Fluke 726

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 726 and is the answer not in the manual?

Fluke 726 Specifications

General IconGeneral
Auto-ramp functionYes
Measuring parametersCurrent, Frequency, Resistance, Temperature, Voltage
Frequency source accuracy0.5 %
Frequency measurement accuracy0.05 %
DC current measurement accuracy0.01 %
DC voltage measurement accuracy0.025 %
Resistance measurement accuracy0.015 %
Product colorBlack, Yellow
Weight and Dimensions IconWeight and Dimensions
Depth47 mm
Width96 mm
Height200 mm
Weight650 g

Related product manuals