Specifications
59
Specifications
Specifications are based on a one year calibration cycle and apply from +18 °C to +28 °C unless stated otherwise. All
specifications assume a five-minute warmup period.
DC Voltage Measurement and Source
Range Minimum Maximum
Accuracy,
(% of Reading + Floor)
30 V (upper display) 0.000 30.000 0.010 % + 2 mV
20 V (lower display) 0.000 20.000 0.010 % + 2 mV
20 V (Source) 0.000 20.000 0.010 % + 2 mV
100 mV (Source) 0.000 100.000 0.010 % + 10 µV
90 mV (Read) 0.000 90.000 0.010 % + 10 µV
Maximum current output in voltage ranges is 1 mA with an output impedance of 1
DC mA Measurement and Source
Range Minimum Maximum
Accuracy,
(% of Reading + Floor)
mA Read (Upper Display) 0.000 24.000
0.010 % + 2 µA
mA Read (Lower Display) 0.000 24.000
0.010 % + 2 µA
mA Source 0.000 24.000
0.010 % + 2 µA
Maximum load on, mA source is 1 ke. With the HART resistor on, maximum load is 750 .
Voltage input range on simulate mode is 5 to 30 V