Table 1. Summary of Source and Measure Functions
Function Measure Source
dc V 0 V to 30 V 0 V to 20 V
dc mA 0 to 24 mA 0 to 24 mA
Frequency 2 CPM to 15 kHz 2 CPM to 15 kHz
Resistance 0 to 4000 5 to 4000
Thermocouple Types E, J, K, T, B, R, S, L, U, N, C, XK, BP
RTD
(Resistance-
Temperature
Detector)
Pt100 (385)
Pt100 (3926)
Pt100 (3916)
Pt200 (385)
Pt500 (385)
Pt1000 (385)
Ni120 (672)
CU10
Pressure 29 modules ranging from 1.0 in. H
2
O to 10,000 psi
Pulse 1-100,000
Freqency Max 10 kHz
1-10,000
Frequency Range 2 CPM to 10 kHz
Other functions Loop supply, HART resistor, pressure switch test, save screen, step, ramp, memory, cold
junction compensation