RTD Accuracy (Read and Source) (ITS-90)
Range Minimum Maximum Accuracy
Ni120 (672) -80.00 260.00 0.15
-200.00 100.00 0.15
100.00 300.00 0.25
300.00 600.00 0.35
Pt100 (385)
600.00 800.00 0.45
-200.00 100.00 0.15
100.00 300.00 0.25
Pt100 (3926)
300.00 630.00 0.35
-200.00 100.00 0.15
100.00 300.00 0.25
Pt100 (3916)
300.00 630.00 0.35
-200.00 100.00 0.75
100.00 300.00 0.85
Pt200 (385)
300.00 630.00 0.95
-200.00 100.00 0.35
100.00 300.00 0.45
Pt500 (385)
300.00 630.00 0.55
-200.00 100.00 0.15
100.00 300.00 0.25
Pt1000 (385)
300.00 630.00 0.35
CU10 -10.00 250.00 1.8
Notes: Read Accuracy is based on 4-wire input. For 3-wire input, add ± 0.05 e assuming all three RTD leads are matched.
Source Accuracy is based on 0.5 to 3.0 mA excitation current (0.1 mA for pt1000 range)