Specifications
61
Frequency Source
Range Minimum Maximum Accuracy
CPM Source 2.0 1000 0.05 %
Hz Source 1.0 1000.0 0.05 %
1.0 10.00 0.25 % KHz Source
10.00 15.00 0.50 %
Temperature, Thermocouples
Type Minimum Maximum
CJC ON
Accuracy
CJC OFF
Accuracy
-210 0.0 0.6 0.4
0.0 800 0.4 0.2
J
800 1200 0.5 0.3
-200 0.0 0.8 0.6
0.0 1000 0.5 0.3
K
1000 1372 0.7 0.5
-250 0.0 0.8 0.6
T
0.0 400 0.4 0.2
-250 -100 0.8 0.6
E
-100 1000 0.4 0.4
-20 0.0 2.0 1.8
R
0.0 1767 1.4 1.2
CJC error outside of 23 ± 5 °C is 0.05 °C / °C