EasyManua.ls Logo

Fluke 726 - Calibrating a Transmitter

Fluke 726
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multifunction Process Calibrator
Calibrating a Transmitter
45
Calibrating a Transmitter
Use the measurement (upper display) and source (lower
display) modes to calibrate a transmitter. This section
applies to all but pressure transmitters. The following
example shows how to calibrate a temperature
transmitter. Use the following steps to calibrate a
transmitter:
1. Connect the Calibrator to the instrument under test
as shown in Figure 20.
2. Press
l for current (upper display). If required,
press
l again to activate loop power.
3. Press
T (lower display). If desired, continue
pressing this key to select the desired thermocouple
type.
4. If necessary, press
M for SOURCE mode.
5. Set the zero and span parameters by pressing
X
and
W. Enter these parameters by pressing and
holding
J and G. For more information on
setting parameters, see “Setting 0 % and 100 %
Output Parameters”.
6. Perform test checks at 0-25-50-75-100 % points by
pressing
H or I. Adjust the transmitter as
necessary.
Note
When HART resistor mode is selected, the
250
resistor is turned on both mA channels.

Table of Contents

Other manuals for Fluke 726

Related product manuals