EasyManuals Logo

Fluke 726 User Manual

Fluke 726
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Multifunction Process Calibrator
Using Measure Mode
25
Table 6. RTD Types Accepted
RTD Type Ice Point (R
0
) Material α Range (°C)
Pt100 (3926) 100 Platinum 0.003926 /°C -200 to 630
Pt100 (385) 100 Platinum 0.00385 /°C -200 to 800
Ni120 (672) 120 Nickel 0.00672 /°C -80 to 260
Pt200 (385) 200 Platinum 0.00385 /°C -200 to 630
Pt500 (385) 500 Platinum 0.00385 /°C -200 to 630
Pt1000 (385) 1000 Platinum 0.00385 /°C -200 to 630
Pt100 (3916) 100 Platinum 0.003916 /°C -200 to 630
The IEC standard RTD commonly used in U.S. industrial applications is the Pt100 (385), α = 0.00385 /°C.
Pt100 (3916), α = 0.003916 /°C is also designated as JIS curve.
Custom RTDs may also be added, see PRT Custom Curves

Table of Contents

Other manuals for Fluke 726

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 726 and is the answer not in the manual?

Fluke 726 Specifications

General IconGeneral
Auto-ramp functionYes
Measuring parametersCurrent, Frequency, Resistance, Temperature, Voltage
Frequency source accuracy0.5 %
Frequency measurement accuracy0.05 %
DC current measurement accuracy0.01 %
DC voltage measurement accuracy0.025 %
Resistance measurement accuracy0.015 %
Product colorBlack, Yellow
Weight and Dimensions IconWeight and Dimensions
Depth47 mm
Width96 mm
Height200 mm
Weight650 g