Table 3-26. Microwave Output Level Accuracy Test, Additional Low Level Test Points, Part A
Part Frequency Amplitude Reading P
out
Tolerance
A
Low Level
Microwave
Output Option
only
10 MHz
-7 dBm Reference Reference
-17 dBm P
rel
±2.0 dB
-27 dBm P
rel
±2.0 dB
-37 dBm (P
-37
) P
rel
±2.0 dB
100 MHz
-7 dBm Reference Reference
-17 dBm P
rel
±2.0 dB
-27 dBm P
rel
±2.0 dB
-37 dBm (P
-37
) P
rel
±2.0 dB
1 GHz
-7 dBm Reference Reference
-17 dBm P
rel
±2.0 dB
-27 dBm P
rel
±2.0 dB
-37 dBm (P
-37
) P
rel
±2.0 dB
4.024 GHz
-7 dBm Reference Reference
-17 dBm P
rel
±2.0 dB
-27 dBm P
rel
±2.0 dB
-37 dBm (P
-37
) P
rel
±2.0 dB
10 GHz
-7 dBm Reference Reference
-17 dBm P
rel
±2.0 dB
-27 dBm P
rel
±2.0 dB
-37 dBm (P
-37
) P
rel
±2.0 dB
20 GHz
-7 dBm Reference Reference
-17 dBm P
rel
±2.0 dB
-27 dBm P
rel
±2.0 dB
-37 dBm (P
-37
) P
rel
±2.0 dB
26.4 GHz
-7 dBm Reference Reference
-17 dBm P
rel
±2.0 dB
-27 dBm P
rel
±2.0 dB
-37 dBm (P
-37
) P
rel
±2.0 dB

Table of Contents