Maintenance
Disassembly and Reassembly 5
5-5
Before You Start
To ensure your safety and for the protection of the Product follow all of the
considerations and recommendations that follow:
 Warning
To prevent possible electrical shock, fire, or personal injury:
Do not remove the covers from the Product unless you are
qualified to do so.
Disconnect all power cords, rear-panel cables, and front/rear
test leads from the Product.
Caution
To prevent possible damage to the product or to equipment
under test:
Removing the internal top cover will void calibration of the
Product.
Removal of the external covers from the Leveling Head will
void its calibration.
Place the Product securely on an appropriate bench top or
work surface before removing the covers. The Product is
heavy.
Use best practice anti-static techniques after removing its
covers.
Do not unnecessarily touch any part of the PCAs or
straighten component positions on the PCAs.
Removing External Hardware Components
Use the subsequent procedures to remove external hardware components from
the Product. Removing all of the hardware components in sequence provides
incremental access to the interior of the Product. See Figure 5-2.
Handles
The Product has four handles, two on the front panel and two on the rear panel.
All four handles are the same. Each handle is secured to the chassis using five
screws, two on the front of the handle and three on the side.
To remove each of the handles:
1. Locate the handle to be removed.
2. Remove the two screws on the front of the handle.
3. Remove the three screws from the side of the handle.