96000 Series
Service Manual
5-12
Note
The display window is secured in place by adhesive tape. Do not
remove the display unless it or the display window is being replaced.
Note
The inside of the front panel has a nickel conductive coat applied to
it. Be careful to avoid damaging the coating.
Digital PCA
The PCA is located behind the front panel and attached to the Product. It can be
removed without removing the top and bottom covers from the Product. See
Figure 5-5.
To remove the Digital PCA:
1. Remove the Front Panel Assembly from the Source.
2. Remove the six panhead screws securing the Digital PCA to the Product.
3. Remove the 60-pin IDE ribbon cable at the top of the PCA by prying the
location lug out of its housing.
4. Remove the 60-pin IDE ribbon cable located on the base of the PCA by
prying the location lug out of the housing.
5. Remove the small 18-way connector.
6. Remove the Digital PCA from the Product.
hpn376.eps
Figure 5-5. Removing the Digital PCA