4-1
Chapter 4
Calibration Adjustment
Introduction
Warning
To prevent possible electrical shock, fire, or personal injury:
Do not perform the verification tests or calibration
procedures described in this manual unless you are
qualified to do so. The information provided in this manual
is for the use of qualified personnel only.
Restrict the humidity of the operating environment to the
level specified for any equipment used that conforms to
IEC60950-1 used with the Product.
Calibration Adjustments
This section of the chapter provides calibration adjustment procedures for
correcting out-of-tolerance parameters so they meet published specifications. If
the Instrument fails the performance test, it is an indication that the Instrument
requires calibration adjustment and/or repair. Calibration adjustment does not
require removal of the covers. See Chapter 5, Maintenance, for internal access
and repair information.
Note
The Instrument top cover is removable for air filter cleaning access
without the need to break calibration integrity seals. A calibration
integrity seal is on the internal top shield and is revealed once the
outer-top cover is removed (see Chapter 5, Maintenance for details).
Additional calibration integrity seals are located on the rear panel
over the lower outer cover and over the calibration enable switch.
Fluke Calibration recommends that users apply calibration integrity
seals in the same three locations.
Calibration adjustment of the Instrument can be performed locally from the
instrument front panel or remotely using the IEEE-488 bus.