Fluke 96000 Series - Phase Noise Test (96040 A and 96270 A, Optional)

Fluke 96000 Series
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
96000 Series
Service Manual
3-12
16. Set the output frequency of the UUT and the Spectrum Analyzer center
frequency to each of the frequencies listed in Table 3-4, repeating steps 14
through 15.
17. Set the UUT output to STBY.
Table 3-4. Spurious Content Test
Output Frequency Limit
2.150 003 GHz < -66 dBc
2.275 003 GHz < -66 dBc
2.425 003 GHz < -66 dBc
2.500 003 GHz < -66 dBc
2.650 003 GHz < -66 dBc
2.825 003 GHz < -66 dBc
2.999 997 GHz < -66 dBc
3.149 997 GHz < -66 dBc
3.424 997 GHz < -66 dBc
3.499 997 GHz < -66 dBc
3.649 997 GHz < -66 dBc
3.774 997 GHz < -66 dBc
3.924 997 GHz < -66 dBc
3.999 997 GHz < -66 dBc
This completes the 96040A harmonics and spurious test points. Additional test
points are required for the 96270A as described in a subsequent procedure. For
the 96040A, leave the connections intact for use in the following tests, if desired.
Phase Noise Test (96040A and 96270A, Optional)
Equipment Required for this Test:
Signal Source Analyzer
50 Ω Leveling Head (supplied with UUT)
Note
This procedure measures at offset frequencies below 1 kHz where
the Signal Source Analyzer models listed in Table 3-1 have only
“Typical” and not guaranteed specification. This should be
considered if any apparent out-of-tolerance results are obtained.
Note
The procedure described below refers to menu selections and
settings for the R&S FSUP Signal Source Analyzer. If using the
alternative Agilent E5052B Signal Source Analyzer, follow the
procedure steps as detailed and substitute settings and menu
selections appropriate to the E5052B to obtain equivalent
measurement conditions.