Performance Tests
Performance Test 3
3-11
11. Repeat steps 5 through 10 for all frequencies in Table 3-3, using the settings
and tolerances shown for each. For UUT output frequency settings of
1.8 GHz and above where the UUT level setting is +14 dBm, set the Analyzer
reference level to +20 dBm.
Table 3-3. Harmonics Test
Level Frequency 2
nd
Harmonic Tolerance 3
rd
Harmonic Tolerance
+24 dBm
20 kHz 40 kHz -60 dBc 60 kHz -60 dBc
2.74 MHz 5.48 MHz -60 dBc 8.22 MHz -60 dBc
5.5 MHz 11 MHz -60 dBc 16.5 MHz -60 dBc
11 MHz 22 MHz -60 dBc 33 MHz -60 dBc
22 MHz 44 MHz -60 dBc 66 MHz -60 dBc
44 MHz 88 MHz -60 dBc 132 MHz -60 dBc
62.5 MHz 125 MHz -60dBc 187.5 MHz -60dBc
88 MHz 176 MHz -60 dBc 264 MHz -60 dBc
125 MHz 250 MHz -60 dBc 375 MHz -60 dBc
+20 dBm
250 MHz 500 MHz -60 dBc 750 MHz -60 dBc
354 MHz 708 MHz -60 dBc 1.062 GHz -60 dBc
500 MHz 1 GHz -60 dBc 1.5 GHz -60 dBc
714 MHz 1.428 GHz -60 dBc 2.142 GHz -60 dBc
1 GHz 2 GHz -60 dBc 3 GHz -60 dBc
1.4 GHz 2.8 GHz -55 dBc 4.2 GHz -55 dBc
+14 dBm
1.8 GHz 3.6 GHz -55 dBc 5.4 GHz -55 dBc
2.7 GHz 5.4 GHz -55 dBc 8.1 GHz -55 dBc
4.024 GHz 8.048 GHz -55 dBc 12.072 GHz -55 dBc
12. Set the Spectrum Analyzer as follows:
PRESET
COUPLING DC
REF LVL +10 dBm
EXT REF On
FREQ 2.150 003 GHz
SPAN 2 MHz
13. Set the UUT as follows:
Mode Leveled Sine
Frequency 2.150 003 GHz
Level +4.6 dBm
Output OPER
14. Set the Spectrum Analyzer to single sweep, initiate a sweep and wait for the
sweep to complete. Push MARKER PEAK.
15. Using the Spectrum Analyzer marker delta functions, measure the amplitude
of any spurious signals detected. Verify any spurious signals at offsets
>±3 kHz from the output frequency within the listed tolerances.