Maintenance
Disassembly and Reassembly 5
5-17
hpn380.eps
Figure 5-9. Removing the Rear Panel
Interconnection PCA
To remove the Interconnection PCA. See Figure 5-9.
1. Remove rear handles.
2. Remove the top and bottom covers from the Product.
3. Remove the internal shield.
4. Remove the RF Output PCA.
5. 96270A - Remove the 27 GHz Control PCA.
6. Remove the Rear Panel Assembly.
7. On the rear panel remove the two serial port connector retaining screw
pillars, the two retaining IEEE 488 connector screw pillars, and the four
panhead screws securing the Interconnection PCA to the rear panel.
8. Lift the Interconnection PCA from the Rear Panel Assembly.