Introduction and Specifications
Specifications 1
1-7
Specifications
General Specifications (96040A and 96270A)
Performance
All specifications apply to a 1-year calibration interval at an ambient temperature
of Tcal ±5 °C.
Nominal factory Tcal calibration temperature 23 °C.
Unless otherwise noted, specifications are stated at a 99 % confidence level
Standard Interfaces
IEEE488.2 (GPIB)
96270A only: Compatible power sensor inputs
[1]
96270A only: USB memory device
Warmup Time 60 minutes
Temperature
Operating: 0 °C to 50 °C
Specified Operation: 5 °C to 40 °C
Storage: -20 °C to +70 °C
Relative Humidity
Operating or Storage:
Non-condensing, 5 °C to 30 °C <90 %, <40 °C <75 %, <70 °C <45 %
Altitude
Operating: ≤2,000 m
Non-operating: ≤12,000 m
Safety EN 61010-1: Pollution Degree 2, Installation Category II
Electromagnetic Compatibility
EN 61326 -1: Controlled.
Applies to use in Korea only. Class A Equipment (Industrial Broadcasting &
Communication Equipment)
[2]
[2] This product meets requirements for industrial (Class A) electromagnetic
wave equipment and the seller or user should take notice of it. This equipment is
intended for use in business environments and is not to be used in homes.
Line Power
Voltage: 100 V to 240 V rms with additional voltage fluctuations of ±10 %.
Frequency: 50 Hz to 60 Hz with additional frequency variation of ±3 Hz.
Power Consumption 250 VA
Dimensions
442 mm (17.4 in) wide, 146 mm (5.76 in) high and 551 mm (21.7 in) deep,
including handles. Mounts within industry-standard 483 mm (19 in) rack-mount
frames when fitted with Y9600 rack mounting kit.
Weight 18 kg (40 lb)
[1] See Power Meter Readout specifications for compatible power sensor details.
442 mm (17.4")
146 mm
(5.75")
133 mm
(5.24")
551 mm (21.7")
hpn366eps
96040A and 96270A Dimensions