5-22
Running Self-Test
The subsequent instructions for the self-test are given in terms of button pushes
from the front panel. The same instructions can also be initiated with the IEEE
488 instructions in a system environment.
Note
The Base level self-test can be run with or without a Leveling Head
attached. However, when a Leveling Head is attached, ensure that
the Leveling Head output is disconnected during self-test and that
there are no earth connections to the floating RF common. This
includes the Leveling Head body and the top two rear BNC
connectors.
To run a self-test:
1. Prepare the Instrument for operation.
2. Push .
3. Push the Support Functions softkey.
4. Push the Self-Test softkey at the bottom of the display. The Self-Test screen
is shown. See Figure 5-12.
The screen shows that no tests have been executed and provides for the
selection of the self-test sequence to run, All, Base, or Head.
hpn85.bmp
Figure 5-12. Select a Self-Test Sequence