Service Manual
5-28
Stage 2
1. Ensure a link has been established between the PC application and the
Product.
The application status box indicates when a connection is established and
the software version of the application. A connection is considered ready
when the line Connection Established.
2. Press Next.
The PC application automatically uploads the necessary firmware
components. A progress bar gives an indication of the status.
The Product also shows a progress bar for erasure and uploading of the
firmware components.
“Upgrade complete: All files transferred successfully” shows on the PC install
screen when the upgrade process is finished.
Stage 3
1. Close the PC application by selecting finish (or pressing the close icon).
2. Restore normal Product operation by setting switch 6 (of the FACTORY SET
dip switches) back in its down position.
3. Power cycle the Product.
The new version of the firmware shows on the start-up screen (and can also be
viewed on the setup screen).