A-2
Frequency Accuracy (96040A and 96270A)
Use a high resolution frequency counter, spectrum analyzer and external
frequency reference to make the measurements shown in Table A-2.
Table A-2. Frequency Accuracy Test
Frequency Actual
Tolerance
(After 24 hour warm-up)
10 kHz ±0.505 mHz
10 MHz ±0.5 Hz
30 MHz ±1.5 Hz
50 MHz ±2.5 Hz
100 MHz ±5.0 Hz
500 MHz ±25.0 Hz
1 GHz ±50.0 Hz
2 GHz ±100.0 Hz
3 GHz ±135.0 Hz
4 GHz ±200.0 Hz