96000 Series
Users Manual
A-18
Table A-18. Attenuation Accuracy (75 Ω) Test (cont.)
Frequency Attenuation Actual Specification Uncertainty Tolerance
100 MHz
0 dB --- Reference Reference Reference
-3 dB ±0.07 dB
-13 dB ±0.07 dB
-23 dB ±0.07 dB
-33 dB ±0.07 dB
-43 dB ±0.1 dB
-53 dB ±0.1 dB
-63 dB ±0.1 dB
-73 dB ±0.2 dB
-82 dB ±0.2 dB
-91 dB ±0.2 dB
-101 dB ±0.4 dB (Typ)
VSWR - 75
Ω
Levelling Head (96040A and 96270A, Optional)
Use a directional bridge and a spectrum analyzer to make the measurements
shown in Table A-19.
Table A-19. VSWR (75 Ω) Test
Level Frequency Actual Tolerance
+7 dBm
500 MHz <1.1
1 GHz <1.2
2 GHz <1.3
-3 dBm
500 MHz <1.1
1 GHz <1.2
2 GHz <1.3
-13 dBm
500 MHz <1.1
1 GHz <1.2
2 GHz <1.3