Performance Test Record
Performance Test Record A
A-21
Level Accuracy - Low Level Microwave Output Option Additional Points (96270A)
Use an AC Measurement Standard, a power meter and power sensors, and a
high performance spectrum analyzer/measuring receiver to make the
measurements shown in Table A-22.
Table A-22. 96270A Low Level Microwave Output Option Additional Level Accuracy Test Points
Level Frequency Actual Typical Tolerance
-7 dBm
1 kHz ±2.0 dB
100 kHz ±2.0 dB
10 MHz ±2.0 dB
100 MHz ±2.0 dB
1 GHz ±2.0 dB
4.024 GHz ±2.0 dB
10 GHz ±2.0 dB
15 GHz ±2.0 dB
20 GHz ±2.0 dB
26.4 GHz ±2.0 dB
-17 dBm
1 kHz ±2.0 dB
100 kHz ±2.0 dB
10 MHz ±2.0 dB
100 MHz ±2.0 dB
1 GHz ±2.0 dB
4.024 GHz ±2.0 dB
10 GHz ±2.0 dB
15 GHz ±2.0 dB
20 GHz ±2.0 dB
26.4 GHz ±2.0 dB
-27 dBm
1 kHz ±2.0 dB
100 kHz ±2.0 dB
10 MHz ±2.0 dB
100 MHz ±2.0 dB
1 GHz ±2.0 dB
4.024 GHz ±2.0 dB
10 GHz ±2.0 dB
15 GHz ±2.0 dB
20 GHz ±2.0 dB
26.4 GHz ±2.0 dB