96000 Series
Service Manual
1-10
Amplitude Flatness relative to
100 kHz
50 Ω Output
Amplitude
dBm
10 Hz
[1]
to
<100 kHz
100 kHz
>100 kHz
to
<10 MHz
10 MHz
to
128 MHz
>128 MHz
to
300 MHz
>300 MHz
to
1.4 GHz
>1.4 GHz
to
3 GHz
>3 GHz
to
4 GHz
>+20 to +24
±0.03 dB ±0.00 dB ±0.04 dB ±0.04 dB
Output not available
>+14 to +20
±0.03 dB ±0.00 dB ±0.04 dB ±0.04 dB ±0.06 dB ±0.2 dB
-17 to +14
±0.03 dB ±0.00 dB ±0.04 dB ±0.04 dB ±0.06 dB ±0.2 dB ±0.3 dB ±0.3 dB
-48 to <-17
±0.03 dB ±0.00 dB ±0.04 dB ±0.04 dB ±0.06 dB ±0.2 dB ±0.3 dB ±0.5 dB
>-74 to <-48
±0.00 dB ±0.2 dB ±0.1 dB ±0.1 dB ±0.4 dB ±0.5 dB ±0.5 dB
>-84 to -74
±0.00 dB ±0.5 dB ±0.1 dB ±0.3 dB ±0.5 dB ±1.0 dB ±1.0 dB
>-94 to -84
±0.00 dB ±0.5 dB ±0.3 dB ±0.5 dB ±1.0 dB ±1.0 dB
-130 to -94 Not Specified
[1] Also applies as typical accuracy at frequencies <10 Hz.
Absolute Amplitude Accuracy
75 Ω Output
Amplitude
dBm
10 Hz
[1]
to
<20 kHz
>20 kHz
to
<100 kHz
>100 kHz
to
<10 MHz
10 MHz
to
125 MHz
>125 MHz
to
300 MHz
>300 MHz
to
1.4 GHz
>1.4 GHz
[2]
to
3 GHz
>3 GHz
[2]
to
4 GHz
>+14 to +18 ±0.12 dB ±0.12 dB ±0.12 dB ±0.12 dB Output not available
>+8 to +14 ±0.12 dB ±0.12 dB ±0.12 dB ±0.12 dB ±0.15 dB ±0.25 dB
-23 to +8 ±0.12 dB ±0.12 dB ±0.12 dB ±0.12 dB ±0.15 dB ±0.25 dB ±0.3 dB ±0.5 dB
-54 to <-23 ±0.15 dB ±0.15 dB ±0.15 dB ±0.15 dB ±0.2 dB ±0.5 dB ±0.5 dB ±0.5 dB
>-80 to <-54
Not Specified
±0.2 dB ±0.2 dB ±0.2 dB ±0.5 dB ±0.5 dB ±0.5 dB
>-90 to -80 ±0.7 dB ±0.7 dB ±0.7 dB ±1.0 dB ±1.0 dB ±1.0 dB
>-100 to -90 ±0.7 dB ±0.7 dB ±0.7 dB ±1.0 dB ±1.0 dB
Not
Specified
-120 to -100 ±1.5 dB ±1.5 dB ±1.5 dB ±1.5 dB
[1] Also applies as typical accuracy at frequencies <10 Hz.