96000 Series
Service Manual
1-12
Microwave Splitter/Sensor Output: Cal Factor Uncertainty
[1] [2]
Calibration Factor uncertainty, applicable to power sensor calibration
Figures in parenthesis
[3]
apply with alternative
user traceability and
correction of mismatch
errors.
100 MHz 1 GHz 2.4 GHz 8 GHz 12 GHz 18 GHz 22 GHz 26.5 GHz
±1.06 %
(±0.37 %)
±1.42 %
(±0.49 %)
±1.42 %
(±0.60 %)
±2.19 %
(±0.76 %)
±2.33 %
(±0.89 %)
±2.91 %
(±1.06 %)
±3.52 %
(±1.36 %)
±3.52 %
(±2.18 %)
[1] Uncertainties are for 95% confidence level (k = 2), and exclude UUT match contribution.
[2] Unless otherwise stated, applies to performance with supplied HF Leveling Kit and sensors, factory calibration and self-
characterization. Self-characterization requires second sensor option.
[3] Calibration factor uncertainty applicable with reference sensor calibrated with typical state-of-the-art uncertainty, and vector
corrections for mismatch errors at splitter output port.
Microwave Splitter/Sensor Output: Power Flatness Uncertainty
[1] [2]
Power Flatness uncertainty, applicable to spectrum analyzer and other measurement device flatness calibration
Power Flatness uncertainty
into 50 Ω (1.0 VSWR
match)
>1 kHz
to
100 MHz
>100 MHz
to
2.4 GHz
>2.4 GHz
to
8 GHz
>8 GHz
to
12.4 GHz
>12.4 GHz
to
18 GHz
>18 GHz
to
26.5 GHz
±0.05 dB ±0.07 dB ±0.10 dB ±0.10 dB ±0.13 dB ±0.16 dB
[1] Uncertainties are specified for 95 % confidence level (k = 2), and exclude UUT match contribution.
[2] Applies to performance with supplied HF Leveling Kit and sensors, factory calibration and self-characterization. Self-
characterization requires second sensor option.
Microwave Splitter/Sensor Output: Voltage Flatness Uncertainty
[1] [2]
Voltage Flatness uncertainty, applicable to oscilloscope bandwidth calibration
Voltage Flatness uncertainty into 50 Ω
>1 kHz
to
100 MHz
>100 MHz
to
2.4 GHz
>2.4 GHz
to
8 GHz
>8 GHz
to
12.4 GHz
>12.4 GHz
to
18 GHz
>18 GHz
to
26.5 GHz
For stated UUT Input VSWR:
1.0
±0.53 % ±0.71 % ±1.10 % ±1.16 % ±1.46 % ±1.76 %
1.2
±1.4 % ±1.49 % ±1.71 % ±1.75 % ±1.96 % ±2.20 %
1.6
±3.38 % ±3.41 % ±3.51 % ±3.53 % ±3.64 % ±3.77 %