EasyManua.ls Logo

Fluke 96000 Series - Page 44

Fluke 96000 Series
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
96000 Series
Service Manual
3-2
Table 3-1. Recommended Equipment
Equipment
Recommended
Model
Minimum Use Requirement
Frequency Counter
Fluke PM 6690
with 2.7 GHz Input option
Period resolution: 100 ns
Frequency resolution: 10 nHz at 10 kHz.
Frequency Standard Fluke 910R
1 x 10
-11
frequency uncertainty
Signal Source
Fluke 9640A, 96040A or
96270A
4 GHz; frequency resolution 10 Hz, short term
amplitude stability 0.1dB
75 Ω Leveling Head required for testing 75 Ω UUTs.
Measuring Receiver/Spectrum
Analyzer
Rohde & Schwarz
FSMR26 with option FS-K40
Frequency resolution: 0.1 Hz
Frequency Accuracy: ± ½ count of lsd
Relative RF Level: ±(0.015 dB ±0.005 dB / 10 dB step)
Frequency: 1 kHz to 4 GHz
Range: +10 to -130 dBm; 50 Ω; Linearity uncertainty,
0.02 dB;
Max. SWR: 1.10; DC coupled;
Phase noise measurement capability: -128 dBc/Hz at
10 kHz and 100 kHz offset at 640 MHz
Signal Source Analyzer
[1]
Rohde & Schwarz FSUP with
option FSUP-B60
or
Agilent E5052B
Phase noise sensitivity, without cross-correlation, at
1 GHz: -108 dBc/Hz at 100 Hz, -144 dBc/Hz at 100
kHz, and -165 dBc/Hz at 10 MHz offset
Phase noise sensitivity, normal capture range, at
1 GHz: -128 dBc/Hz at 1 kHz, -144 dBc/Hz at 100
kHz, and -170 dBc/Hz at 10 MHz offset
AC Measurement Standard Fluke 5790A
Accuracy: ± 0.1%, 200 Hz - 100 kHz
Precision Type-N, Female-to-Female
Adapter, 50 Ω
Maury Microwave 8828A
VSWR: DC to 4 GHz, 1.03; Insertion loss characterized:
100 kHz to 4 GHz
Precision Type-N, Female-to-Female
Adapter, 75 Ω
Maury Microwave 8882A
VSWR: DC to 2 GHz, 1.03; Insertion loss characterized,
100 kHz to 2 GHz
50 Ω Type-N Female-to-BNC Male
Adapter
Maury Microwave 8821B1
VSWR: DC to 4 GHz, 1.08
75 Ω Type-N Female-to-BNC Male
Adapter
Agilent 1250-1534
VSWR: DC to 3 GHz, 1.03
50 Ω Directional Bridge
Agilent 86205A
Directivity: 40 dB, 5 MHz to 2 GHz
75 Ω Directional Bridge
Agilent 86207A
Directivity: 40 dB, 5 MHz to 1.3 GHz
Attenuator, 6 dB, 50 Ω, N-Type
Weinschel Model 1-6
Male/female. VSWR: DC to 4 GHz, 1.15
Attenuator, 6 dB, 50 Ω, 2.92mm
[2]
Anritsu Model 41KB-6
Male/female. VSWR: DC to 26.5 GHz, 1.3
Precision 50 Ω Open/Short Termination
Narda Microwave 231-416
Minimum Reflection Coefficient: 0.99
Precision 75 Ω Open Termination
Maury Microwave 8885B
Minimum Reflection Coefficient: 0.98
Precision 75 Ω Short Termination
Maury Microwave 8884B
DC – 2 GHz
Precision 50 Ω Feedthrough
Termination
[3]
50 Ω ± 0.2%
Precision 75 Ω Feedthrough
Termination
[4]
75 Ω ± 0.2%
Impedance Matching Pad, 75 / 50 Ω
Rohde & Schwarz RAZ
75 Ω Type-N (f) – 50 Ω Type-N (m)
Frequency range: DC to 2.7 GHz
Insertion loss: 1.76 dB
VSWR: 0 to 2 GHz, 1.06; 2 to 2.7 GHz, 1.2
Flatness: < 0.025 dB, DC – 0.5 GHz
Power Meter Rohde & Schwarz NRVD
Compatible with power sensors
Power Meter
[5]
Rohde & Schwarz NRP2
Compatible with power sensors
Power Sensor, Thermal, 50 Ω
Rohde & Schwarz NRV-Z51
or
Rohde & Schwarz NRP-Z51
Type-N Connector
Frequency: 100 kHz to 4 GHz
Power: +20 to -20 dBm
Cal factor uncertainty: ±0.3%;
Max. SWR: 1.10; DC coupled

Table of Contents

Related product manuals