96000 Series
Service Manual
3-2
Table 3-1. Recommended Equipment
Equipment
Recommended
Model
Minimum Use Requirement
Frequency Counter
Fluke PM 6690
with 2.7 GHz Input option
Period resolution: 100 ns
Frequency resolution: 10 nHz at 10 kHz.
Frequency Standard Fluke 910R
1 x 10
-11
frequency uncertainty
Signal Source
Fluke 9640A, 96040A or
96270A
4 GHz; frequency resolution 10 Hz, short term
amplitude stability 0.1dB
75 Ω Leveling Head required for testing 75 Ω UUTs.
Measuring Receiver/Spectrum
Analyzer
Rohde & Schwarz
FSMR26 with option FS-K40
Frequency resolution: 0.1 Hz
Frequency Accuracy: ± ½ count of lsd
Relative RF Level: ±(0.015 dB ±0.005 dB / 10 dB step)
Frequency: 1 kHz to 4 GHz
Range: +10 to -130 dBm; 50 Ω; Linearity uncertainty,
0.02 dB;
Max. SWR: 1.10; DC coupled;
Phase noise measurement capability: -128 dBc/Hz at
10 kHz and 100 kHz offset at 640 MHz
Signal Source Analyzer
[1]
Rohde & Schwarz FSUP with
option FSUP-B60
or
Agilent E5052B
Phase noise sensitivity, without cross-correlation, at
1 GHz: -108 dBc/Hz at 100 Hz, -144 dBc/Hz at 100
kHz, and -165 dBc/Hz at 10 MHz offset
Phase noise sensitivity, normal capture range, at
1 GHz: -128 dBc/Hz at 1 kHz, -144 dBc/Hz at 100
kHz, and -170 dBc/Hz at 10 MHz offset
AC Measurement Standard Fluke 5790A
Accuracy: ± 0.1%, 200 Hz - 100 kHz
Precision Type-N, Female-to-Female
Adapter, 50 Ω
Maury Microwave 8828A
VSWR: DC to 4 GHz, ≤ 1.03; Insertion loss characterized:
100 kHz to 4 GHz
Precision Type-N, Female-to-Female
Adapter, 75 Ω
Maury Microwave 8882A
VSWR: DC to 2 GHz, ≤ 1.03; Insertion loss characterized,
100 kHz to 2 GHz
50 Ω Type-N Female-to-BNC Male
Adapter
Maury Microwave 8821B1
VSWR: DC to 4 GHz, ≤ 1.08
75 Ω Type-N Female-to-BNC Male
Adapter
Agilent 1250-1534
VSWR: DC to 3 GHz, ≤ 1.03
50 Ω Directional Bridge
Agilent 86205A
Directivity: 40 dB, 5 MHz to 2 GHz
75 Ω Directional Bridge
Agilent 86207A
Directivity: 40 dB, 5 MHz to 1.3 GHz
Attenuator, 6 dB, 50 Ω, N-Type
Weinschel Model 1-6
Male/female. VSWR: DC to 4 GHz, ≤ 1.15
Attenuator, 6 dB, 50 Ω, 2.92mm
[2]
Anritsu Model 41KB-6
Male/female. VSWR: DC to 26.5 GHz, ≤ 1.3
Precision 50 Ω Open/Short Termination
Narda Microwave 231-416
Minimum Reflection Coefficient: 0.99
Precision 75 Ω Open Termination
Maury Microwave 8885B
Minimum Reflection Coefficient: 0.98
Precision 75 Ω Short Termination
Maury Microwave 8884B
DC – 2 GHz
Precision 50 Ω Feedthrough
Termination
[3]
50 Ω ± 0.2%
Precision 75 Ω Feedthrough
Termination
[4]
75 Ω ± 0.2%
Impedance Matching Pad, 75 / 50 Ω
Rohde & Schwarz RAZ
75 Ω Type-N (f) – 50 Ω Type-N (m)
Frequency range: DC to 2.7 GHz
Insertion loss: 1.76 dB
VSWR: 0 to 2 GHz, ≤ 1.06; 2 to 2.7 GHz, ≤ 1.2
Flatness: < 0.025 dB, DC – 0.5 GHz
Power Meter Rohde & Schwarz NRVD
Compatible with power sensors
Power Meter
[5]
Rohde & Schwarz NRP2
Compatible with power sensors
Power Sensor, Thermal, 50 Ω
Rohde & Schwarz NRV-Z51
or
Rohde & Schwarz NRP-Z51
Type-N Connector
Frequency: 100 kHz to 4 GHz
Power: +20 to -20 dBm
Cal factor uncertainty: ±0.3%;
Max. SWR: 1.10; DC coupled