3-48
32. Return the UUT output to -81dBm, set the UUT frequency to the next
frequency listed in part A of Table 3-21, and repeat steps 29 through 32.
33. Set the UUT to 10 MHz at -91dBm. Repeat steps 39 through 43 for the points
listed in part B of Table 3-21 using the calculation P
out
= P
-91
+ P
rel
in step 30.
34. Set the UUT to STBY.
Table 3-21. 75 Ω Leveling Head Optional Ultra-Low Level Accuracy Test (96040A and 96270A)
Part Frequency Amplitude Reading P
out
Tolerance
A
10 MHz
-81 dBm Reference Reference
-101dBm P
rel
±1.50 dB
-121dBm P
rel
±1.50 dB
100 MHz
-81 dBm Reference Reference
-101dBm P
rel
±1.50 dB
-121dBm P
rel
±1.50 dB
300 MHz
-81 dBm Reference Reference
-101dBm P
rel
±1.50 dB
-121dBm P
rel
±1.50 dB
1.4 GHz
-81 dBm Reference Reference
-101dBm P
rel
±1.50 dB
-121dBm P
rel
±1.50 dB
2 GHz
-81 dBm Reference Reference
-101dBm P
rel
±1.50 dB
-121dBm P
rel
±1.50 dB
B
10 MHz
-91 dBm Reference Reference
-111 dBm P
rel
±1.50 dB
100 MHz
-91 dBm Reference Reference
-111 dBm P
rel
±1.50 dB
300 MHz
-91 dBm Reference Reference
-111 dBm P
rel
±1.50 dB
1.4 GHz
-91 dBm Reference Reference
-111 dBm P
rel
±1.50 dB
2 GHz
-91 dBm Reference Reference
-111 dBm P
rel
±1.50 dB