EasyManua.ls Logo

Fluke PM6685 - Power Supply Calibration

Fluke PM6685
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction
Required Test Equipment
* For adjustment of PM9691 and PM9692 Oven Oscillators only.
Note: Only calibrated instruments should be used.
Preparation
WARNING: Live parts and accessible terminals which
can be dangerous to life are always exposed inside
the unit when it is connected to line power. Use ex
-
treme caution when handling, testing, or adjusting
the counter.
Before beginning the calibration adjustments, power up the instru
-
ment and leave it on for at least 30 minutes to let it reach normal oper
-
ating temperature.
Power Supply
CAUTION: If you adjust the +5 V trimmer you have to
adjust the complete instrument.
n
Setup
Remove the protective cover above the power module.
WARNING: The heat sink inside the power module is
connected to line power.
Connect the counter to line power.
Switch on the counter.
Press PRESET, then press ENTER.
NOTE: The backlight must be switched on during the adjust
-
ment of the power module.
n
Adjustment
Connect the DMM to test point TP15 = +5V and GND, (see
Fig. 6-2).
Adjust the +5V trim potentiometer inside the power module
until the DMM reads +5.10 ± 0.01 V.
Check that the voltage between the test point TP23 = +5 V and
GND is +5.06 ± 0.03 V.
Check that the unregulated voltage from the power module at
test point TP16 = +15 V is about +18 V.
Check that the unregulated voltage from the power module at
test point TP17 = –7 V is about –8 V.
Reinstall the protective cover onto the power module.
6-2 Introduction
Power
Module
Fuse
TP23, +5
TP20, -5.2
TP22, +7
J10
J9
J4
J3
+5V adjust
6 14
1 5
TP21,+12V
TP15, +5
TP17, -7
J15
TP16, +15
Load
Disconnected
Connected
J21
J16
Fig. 6-1 Test points and trimmer for the Power Supply.
Type Performance
DMM Acc. 0.02% / Res. 1mV
HF synthesizer 3300 MHz
Pulse generator 125 MHz/2nsrise/fall time
LF synthesizer 50 MHz / 20 Vpp
Oscilloscope 300 MHz / 2-channel
Passive probe 10:1, preferably 500 W (or well
compensated 10 MW)
FET probe 300 MHz
Power supply 12V/2A
Power splitter 50 W /4W
Feed-through termination 50 W
10 MHz reference 1x10
-7
10 MHz reference 1x10
-9
*
BNC-BNC cables Different lengths
Screwdrivers Torx 10 & 20
Table 6-1 Required test equipment.

Other manuals for Fluke PM6685