Display Hold: Freezes meas. result until
a new measurement is ini
-
tiated via Restart.
Limit Alarm:
Limit Values:
Settings:
On Alarm:
Display:
Numeric:
Graphic:
Annunciator on display
and/or SRQ via GPIB
Lower limit (limit 1)
Upper limit (limit 2)
Off
Alarm if value > limit 2
Alarm if value < limit 1
Alarm if
limit 1 < value < limit 2
Alarm if value > limit 2 or
value < limit 1
Stop or Continue
Numeric + graphic
Current measurement
value + annunciator for
limit alarm
Horizontal line w. up-
per/lower limit markers +
current value marker
Stored Instrument
Setups: 20 complete setups can
be saved and recalled
from internal non-volatile
memory; 10 can be user
protected
Display
Type & Use:
Resolution:
Technology:
Graphics screen for menu
control, numerical readout
(14 digits) and status info
320 x 97 pixels
Monochrome LCD with
white LED backlight
GPIB Interface
Programmable
Functions: All front panel accessible
functions
Compatibility: IEEE 488.2-1987,
SCPI 1999
Modes: Native mode
Agilent compatible mode
Agilent
Compatibility: HP 53131/132/181 com
-
mands are emulated.
Code and response for
-
mat is compatible. No tim
-
ing compatibility. No reso
-
lution compatibility
Interface Func
-
tions: SH1, AH1, T6, L4, SR1,
RL1, DC1, DT1, E2
Max. Meas. Rate
Via GPIB:
To Internal
Memory:
2000 readings/s (block)
350 readings/s (individual)
250 k readings/s
Internal Memory
Size: Up to 750 k readings
Data Output: ASCII, IEEE double preci-
sion floating point
USB Interface
Version: 2.0, 12 Mb/s
Protocol: USBTMC-USB488
TimeViewâ„¢
This software package is intended for ad
-
vanced Time & Frequency analysis and runs
on any 32-bit Windows
®
system.
n
Data Capture Modes &
Measurement Rate
Effective rate depends on measurement func
-
tion and internal data format.
Free-running
Measurement: 250 k readings/s
Specifications
8-8 Auxiliary Functions