Activate start arming by keying in the fol
-
lowing sequence:
SETTINGS ® Start Chan ® E
The counter does not measure.
Apply one single pulse to Ext Arm Input.
The counter measures once and shows
10 MHz on the display.
Measuring
Functions
Connect a 10 MHz sine wave signal with
approx. 1 V
RMS
amplitude into 50 W via a
power splitter to Input A and Input B, e.g.
from 10 MHz Out on the rear panel. Use
equal cable lengths from the power splitter
to the inputs.
Recall the DEFAULT settings.
Select the following settings for the
timer/counter via INPUT A and INPUT B:
50 W impedance for A and B.
MANual trigger.
POS slope.
Check that the timer/counter performs the
correct measurement by displaying the re
-
sult as shown under the “Display” column
in Table 7-7.
Select function via MEAS FUNC.
Check of HOLD
OFF Function
Recall the DEFAULT settings.
Select the following common timer/counter
settings for both Input A and Input B via the
hard menu keys INPUT A and INPUT B:
50 W impedance.
DC coupling.
MANual trigger, x1 attenuation.
Trigger level = 0.5 V.
Press SETTINGS and activate Hold Off.
Select Hold Off On and set the Trigger
Hold Off time to 1 ms.
Connect the pulse generator with the fol-
lowing settings to Input A :
Period = 100 ms.
Duration 10 ns.
Double pulse.
Delay = 1 ms.
Amplitude = 1.0 V
PP
, (high level +1V and
low level 0V).
Rise time 2 ns.
Check the following results:
Freq A with Hold Off On = 10 kHz.
Freq A with Hold Off Off = 20 kHz.
Connect the signal to Input B, select
Freq B and repeat the tests above.
Performance Check
7-10 Measuring Functions