Calibration
Mode: Closed case, menu-con
-
trolled.
Calibration Input: A
Password
Protection: ON or OFF
Input Frequen
-
cies used for TB
Calibration: 1.0, 1.544, 2.048, 5.0 or
10.0 MHz
Definition of Terms
Calibration Ad
-
justment Toler-
ance: The maximum tolerated
deviation from the true 10
MHz frequency after cali-
bration. If the timebase
frequency does not ex-
ceed the tolerance limits
at the moment of calibra-
tion, an adjustment is not
necessary.
Total Uncertainty: The total possible devia
-
tion from the true 10 MHz
frequency under influence
of frequency drift due to
aging and ambient tem
-
perature variations versus
the reference tempera
-
ture. The operating tem
-
perature range and the
calibration interval are
part of this specification.
General
Specifications
Environmental Data
Class: MIL-PRF-28800F, Class 3
Operat. Temp:
Storage Temp:
Humidity:
Max. Altitude:
Vibration:
Shock:
Transit Drop
Test:
0 °C to +50 °C
-40 °C to +71 °C
5-95 % @ 10-30 °C
5-75 % @ 30-40 °C
5-45 % @ 40-50 °C
4600 m
Random and sinusoidal per
MIL-PRF-28800F, Class 3
Half-sine 30G per
MIL-PRF-28800F; bench
handling
Transport box tested ac-
cording to UN-D 1400 drop
test program 1.
Heavy-duty transport case
and soft carrying case
tested according to
MIL-PRF-28800F.
Reliability:
Safety:
MTBF 30000 h, calculated
Designed and tested for
Measurement Category I,
Pollution Degree 2, in ac
-
cordance with
EN/IEC 61010-1:2001 and
CAN/CSA-C22.2 No.
61010-1-04 (incl. approval)
EMC: EN 61326 (1997)
A1 (1998), increased test
levels per EN 50082-2,
Group 1, Class B, CE
Power Requirements
Line Voltage:
Power
Consumption:
90-265 V
RMS
, 45-440 Hz
<40 W
Basic version w/o options.
Calibration 8-11
Specifications